Diagnostics and troubleshooting
Smartec CLD18
36
Hauser
Diagnostic code
Message text
Description
F523
Cell constant
Sensor calibration warning
Invalid cell constant, max. range reached
Remedy:
• Enter cell constant as per factory specifications
• Contact the Service Department
F524
Cell constant
Sensor calibration warning
Min. possible cell constant is undershot
Remedy:
• Enter cell constant as per factory specifications
• Contact the Service Department
F845
Device id
Incorrect hardware configuration
F847
Couldn’t save param
Incorrect parameters
F848
Calib AO1
Incorrect calibration values for analog output 1
F849
Calib AO2
Incorrect calibration values for analog output 2
F904
Process check
Process check system alarm
Measuring signal has not changed for a long time
Possible reasons:
• Contaminated sensor, or sensor in air
• No flow to sensor
• Sensor defective
• Software error
Remedy:
• Check electrode system
• Check sensor
• Restart device
Diagnostic code
Message text
Description
C107
Calib. active
Sensor calibration is active
Remedy:
Wait for calibration to be finished
C154
No calib. data
Sensor data
No calibration data available, factory settings are used
Remedy:
• Check the calibration information of the sensor
• Contact the Service Department
C850
Simu AO1
Simulation of analog output 1 is active
C851
Simu AO2
Simulation of analog output 2 is active