Smartec CLD18
Electrical connection
Hauser
19
-
+
L
L
L
+
L
-
O
u
t+
O
u
t-
MR
S+
MR
S-
I
u
t1
-
O
I
u
t2
+
O
I
u
t1
+
O
I
u
t2
-
O
X
1
2
A0029684
10
Terminal assignment
IOut1
Current output conductivity (active)
IOut2
Current output temperature (active)
Out
Alarm output (open-collector)
MRS
Binary input (measuring range switch)
L+/L-
Power supply
X
Grounding pin (flat male tab 4.8 mm)
1
Cover on electronics box
2
Electronics box
NOTICE
Removing the electronics box will destroy the sensor connection!
‣
The electronics box must not be removed under any circumstances.
‣
Do not open the cover on the electronics box.
The recommended cable cross-section for the connecting cables is 0.5 mm
2
. The
maximum cable cross-section is 1.0 mm
2
.
Connect the transmitter of the compact device as follows:
1. Unscrew the housing cover.