Electrical connection
Smartec CLD18
20
Hauser
2. Guide the connecting cables through the cable glands.
3. Connect the cables as per the terminal assignment diagram.
4. Connect the protective ground to the terminal pin for the housing ground.
6.1.2
Connection via M12 connector
3
2
1
4
A0033108
11
View of connector, 4-pin, data cable (at device)
1
IOUT1+
Conductivity
3
IOUT2-
Temperature
2
IOUT2+
Temperature
4
IOUT1-
Conductivity
4
3
2
1
7
6
5
8
A0033109
12
View of connector, 8-pin, power supply/controller (at device)
1
L+
Power supply
5
Out+
Alarm
2
L-
Power supply
6
Out-
Alarm output-
3
MRS+
Binary input
7
GND
Functional ground
4
MRS-
Binary input
8
GND
Functional ground
Содержание Smartec CLD18