Commissioning
Smartec CLD18
32
Hauser
Calibrating the cell constant
‣
With this type of calibration, enter a reference value for the conductivity.
In the result, the device calculates a new cell constant for the sensor.
First switch off the temperature compensation:
1. Select the menu
Setup
/
Extended setup
/
Input
/
Temp. comp.
.
2.
Off
Select .
3. Return to the menu
Setup
.
Perform the calculation of the cell constant as follows:
1. Select the menu
Calibration
/
Cell const.
.
2.
Cond. ref.
Select and enter the value of the standard solution.
3. Place the sensor in the medium.
4. Start the calibration.
"
Wait calib.
" - wait for calibration to end. The new value is displayed after the
calibration.
5. Press the Plus key.
"
Save calib data?
"
6.
Yes
Select .
"
Calib successful
"
7. Switch the temperature compensation back on.