Smartec CLD18
Diagnostics and troubleshooting
Hauser
37
Diagnostic code
Message text
Description
S844
Process value
Measured value outside the specified range
Possible reasons:
• Sensor in air
• Incorrect flow to sensor
• Sensor defective
Remedy:
• Increase process value
• Check electrode system
Diagnostic code
Message text
Description
M500
Not stable
Sensor calibration aborted
Main measured value fluctuating
Possible reasons:
• Sensor in air
• Sensor fouled
• Incorrect flow to sensor
• Sensor defective
Remedy:
• Check sensor
• Check installation
M526
Cell constant
Sensor calibration warning
Invalid cell constant, max. range reached
Remedy:
• Repeat the calibration
• Enter cell constant as per factory specifications
• Contact the Service Department
M528
Cell constant
Sensor calibration warning
Min. possible cell constant is undershot
Remedy:
• Repeat the calibration
• Enter cell constant as per factory specifications
• Contact the Service Department