Electrical connection
Proline Promag L 400 HART
32
Hauser
Coil current cable
Standard cable
2
´
0.75 mm
2
(18 AWG) with common, braided copper shield (
Æ
~
7 mm (0.28")) and
individually shielded cores
Conductor resistance
£
37
W
/km (0.011
W
/ft)
Capacitance: core/core,
shield grounded
£
120 pF/m (37 pF/ft)
Operating temperature
–20 to +80 °C (–68 to +176 °F)
Test voltage for cable
insulation
£
AC 1433 V r.m.s. 50/60 Hz or
³
DC 2026 V
1
2
3
4
5
6
7
a
b
A0003194
å
8
Cable cross-section
a
Electrode cable
b
Coil current cable
1
Core
2
Core insulation
3
Core shield
4
Core jacket
5
Core reinforcement
6
Cable shield
7
Outer jacket
Reinforced connecting cable
Use in the following situations:
• When laying the cable directly in the ground
• Where there is a risk of damage from rodents
Operation in zones of severe electrical interference
The measuring system meets the general safety requirements (
Grounding is by means of the ground terminal provided for the purpose inside the connection
housing. The stripped and twisted lengths of cable shield to the ground terminal must be as short
as possible.
Cable diameter
• Cable glands supplied:
– For standard cable: M20 × 1.5 with cable
Æ
6 to 12 mm (0.24 to 0.47 in)
– For reinforced cable: M20 × 1.5 with cable
Æ
9.5 to 16 mm (0.37 to 0.63 in)
• (Plug-in) screw terminals for wire cross-sections 0.5 to 2.5 mm
2
(20 to 14 AWG)