Proline Promag L 400 HART
Technical data
Hauser
125
Input signal
Status input
Maximum input values
• DC 30 V
• 6 mA
Response time
Adjustable: 5 to 200 ms
Input signal level
• Low signal: DC –3 to +5 V
• High signal: DC 12 to 30 V
Assignable functions
• Off
• Reset totalizers 1-3 separately
• Reset all totalizers
• Flow override
16.4
Output
Output signal
Current output
Current output
Can be set as:
• 4-20 mA HART (active)
• 0-20 mA (active)
Maximum output values
• DC 24 V (when idle)
• 22.5 mA
Load
0 to 700 Ω
Resolution
0.5 µA
Damping
Adjustable: 0 to 999 s
Assignable measured
variables
• Volume flow
• Conductivity
• Mass flow
Pulse/frequency/switch output
Function
• With the order code for "Output; Input", option
H
: output 2 can be set as a pulse or
frequency output
• With the order code for "Output; Input", option
I
: output 2 and 3 can be set as a pulse,
frequency or switch output
Version
Passive, open collector
Maximum input values
• DC 30 V
• 250 mA
Voltage drop
For 25 mA:
£
DC 2 V
Pulse output
Pulse width
Adjustable:0.05 to 2
000 ms
Maximum pulse rate
10
000 Impulse/s
Pulse value
Adjustable
Assignable measured
variables
• Volume flow
• Mass flow
Frequency output
Output frequency
Adjustable: 0 to 10
000 Hz
Damping
Adjustable: 0 to 999 s
Pulse/pause ratio
1:1