You can assign the following options to the diagnostic number as the diagnostic behavior:
Options
Description
Alarm
Measurement is interrupted. Signal outputs and totalizers assume the defined alarm condition. A
diagnostic message is generated.
The background lighting changes to red.
Warning
Measurement is resumed. The signal outputs and totalizers are not affected. A diagnostic message
is generated.
Logbook entry only
The device continues to measure. The diagnostic message is entered in the Event logbook (events
list) submenu only and is not displayed in alternation with the measured value display.
Off
The diagnostic event is ignored, and no diagnostic message is generated or entered.
12.6
Overview of diagnostic information
Diagnostic
number
Short text
Remedy information
Status
signal
from the
factory
Diagnostic behavior
from the factory
Diagnostics for the sensor
043
Sensor short-circuit
1. Check the sensor and cable.
2. Replace the sensor or cable.
S
Warning
*
062
Sensor connection
1. Check the sensor connections.
2. Contact service.
F
Alarm
*
Diagnostic behavior can be changed: "Adapting the diagnostic behavior" section (
Diagnostic
number
Short text
Remedy information
Status
signal
from the
factory
Diagnostic behavior
from the factory
Diagnostics for the electronics
242
Software incompatible
1. Check software.
2. Flash or change main electronic
module.
F
Alarm
270
Main electronic failure
Change main electronic module.
F
Alarm
271
Main electronic failure
1. Restart device.
2. Change main electronic module.
F
Alarm
272
Main electronic failure
1. Restart device.
2. Contact service.
F
Alarm
273
Main electronic failure
Replace electronics.
F
Alarm
283
Memory content
1. Reset device.
2. Contact service.
F
Alarm