Proline Promag L 400 HART
Electrical connection
Hauser
41
7.4
Post-connection check
Are cables or the device undamaged (visual inspection)?
q
Do the cables comply with the requirements (
q
Do the cables have adequate strain relief?
q
Are all the cable glands installed, firmly tightened and leak-tight? Cable run with "water trap" (
q
Only for remote version: is the sensor connected to the right transmitter?
Check the serial number on the nameplate of the sensor and transmitter.
q
Does the supply voltage match the specifications on the transmitter nameplate (
q
Is the terminal assignment correct (
q
If supply voltage is present, do values appear on the display module?
q
Is the potential equalization established correctly (
q
Are all housing covers installed and the screws tightened with the correct tightening torque?
q