Function group PARAMETER
Verification of the amplifier takes place along four measuring points MP1 to MP4. Enter the full
scale value MP4 of the verification flow (100%) and the two selectable measuring points MP2 and
MP3 (in % of the full scale value) for the STANDARD RANGE option in the FLOW RANGE function.
!
Note!
The value of MP1 is prespecified (5% of the full scale value MP4).
Assign the required flow values to measuring points MP1 to MP4 for the SPEC. APPLICATION
option in the FLOW RANGE function.
Function description
FUNCTION
VERIFICATION
PARAMETER
FLOW 100%
Use this function to enter the full scale value of the verification flow.
User input:
Max 6-digit floating point number (without sign)
!
Note!
The permitted value range depends on the value set in the FLOW RANGE function
67.
The input limits MIN and MAX are displayed below the input line.
Unit:
The unit configured in the measuring device is adopted.
Factory setting:
Equivalent to 4 m/s (depending on the unit)
MEAS. POINT 2/3
Use this function to enter the measuring points MP2 and MP3 (in % of the full scale
value).
User input:
MP2: 10 to 90 %
MP3: 15 to 95 %
!
Note!
• The value of MP2 must be less than the value of MP3.
• Both values are stored permanently.
Factory setting:
MP2: 10 %
MP3: 50%
MEASURING POINT
MP1 to MP4
If you have selected the SPEC. APPLICATION option in the FLOW RANGE function,
enter the flow values for the four measuring points here.
The MIN and MAX limit values are displayed. The values entered for the flows for MP1
to MP4 must be higher than the value for the LOW FLOW CUT OFF - DEACTIVATION
measuring device parameter.
User input:
Max 6-digit floating point number (without sign)
Unit:
The unit configured in the measuring device is adopted.
Factory setting:
MP1: 0.0 (unit)
MP2: 0.0 (unit)
MP3: 0.0 (unit)
MP4: 0.0 (unit)