Proline Fieldcheck
Device functions - general section
Hauser
47
RESULT
VERIFICATION
SENSOR MID
The measured values are grouped and displayed over a maximum of 5 lines.
Use the ENTER key to go to the next group.
Display:
Measurement of ground connection, optional: measurement of magnetic coil resistance.
Measurement of electrode resistance (only if the measuring tube is completely full).
Isolation measurement, optional: HV isolation measurement
RESULT
VERIFICATION
TRANSMITTER
CORIOLIS
Use this function to call up the results of the verification. The result (PASSED, FAILED) is
displayed on the information line. If the results of a linearity check are called up, the list
of measured values also appears when confirming the selection.
Options:
ZERO POINT
AMPLIFIER (with list of measured values)
DENSITY (with list of measured values)
TEMPERATURE (with list of measured values)
CURRENT OUTPUT (with list of measured values)
FREQUENCY OUTPUT (with list of measured values)
PULSE MEASUREMENT
SENSOR
RESULT
VERIFICATION
SENSOR CORIOLIS
The measured values are grouped and displayed over a maximum of 5 lines.
Use the ENTER key to go to the next group.
Display:
Measurement of the ground connection
Measurement of excitation coil resistance
Measurement of signal coil resistance
Measurement of temperature sensor resistance
Isolation measurement, optional: HV isolation measurement
RESULT
VERIFICATION
VORTEX
Use this function to call up the results of the verification. The result (PASSED, FAILED) is
displayed on the information line. If the results of a linearity check are called up, the list
of measured values also appears when confirming the selection.
Options:
ZERO POINT
AMPLIFIER (with list of measured values)
CURRENT OUTPUT (with list of measured values)
P-73 (temperature, with list of measured values)
PULSE MEASUREMENT
SENSOR
RESULT
VERIFICATION
ULTRASONIC
Use this function to call up the results of the verification. The result (PASSED, FAILED) is
displayed on the information line. If the results of a linearity check are called up, the list
of measured values also appears when confirming the selection.
Options:
ZERO POINT
AMPLIFIER (with list of measured values)
CURRENT OUTPUT (with list of measured values)
FREQ. OUTPUT (with list of measured values)
PULSE MEASUREMENT
SENSOR (only clamp-on sensors)
RESULT
VERIFICATION
T-MASS
Use this function to call up the results of the verification. The result (PASSED, FAILED) is
displayed on the information line.
If the results of a linearity check are called up, the list of measured values also appears
when confirming the selection.
Options:
POWER (with list of measured values)
RESISTANCE (with list of measured values)
FLOW (with list of measured values)
CURRENT (with list of measured values)
FREQUENCY (with list of measured values)
PULSE
SENSOR
Function description
SUPERVISION
RESULTS