Proline Fieldcheck
Wiring
Hauser
17
3.5
Connecting to Promass 40, 80, 83, 84 field housing
"
Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the connection compartment.
1.
Remove the cover of the electronics compartment (5).
2.
Remove the local display.
3.
Disconnect the excitation current cable from the excitation current socket (6).
#
Warning!
The excitation current cable must be disconnected as otherwise the measuring tube could
vibrate in the resonance frequency and become damaged!
4.
Pull the sensor signal cable connector out of the amplifier socket (7).
5.
Plug the Simubox cable connector into the amplifier socket (7).
The adapter (
11) should be used to connect the Simubox to Promass 40 measuring
devices.
6.
Remove the cover of the connection compartment (8) from the transmitter housing.
7.
Connect Fieldcheck to the current and frequency outputs of the transmitter
(terminal pairs 22/23, 24/25 or 26/27).
!
Note!
The configuration of the outputs depends on the product type and the order version of the
tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
"Wiring" section of the flowmeter Operating Instructions.
8.
Connect Fieldcheck to the service connector (9).
a0001700
Fig. 10:
Connecting Fieldcheck to Promass 40, 80, 83, 84 (field housing)
1
Fieldcheck flow simulator
2
Simubox
3
Stainless steel transmitter
4
Aluminum transmitter
5
Electronics compartment cover
6
Excitation current socket
7
Amplifier socket
8
Connection compartment cover
9
Service connector
27
25
23
21
2
1
26
24
22
20
L1 (L+)
N (L-)
8
3
5
4
8
5
7
6
2
1
9