Proline Fieldcheck
Device functions - electromagnetic measuring systems
Hauser
51
FLOW SPECIFICATION
Use this function to select which measurement is to be simulated.
Options:
VOLUME FLOW
MASS FLOW
!
Note!
• "Calculated mass flow" simulation is only possible with Promag 53 type devices.
• The value stored in the DENSITY VALUE function (DENSITY PARAMETER function
group of the device) is used for the calculation.
• If a current input is available in the device and the density is assigned to it, mass flow
simulation is
not
possible.
Factory setting:
VOLUME FLOW
SIGNAL 1 LEVEL/
SIGNAL 2 LEVEL
Use this function to enter the flow value which is to be simulated.
Enter the second flow value which is to be simulated in SIGNAL 2 LEVEL if you have
selected the STEP or PROFILE operating mode.
!
Note!
More information is provided in the OPERATION MODE function.
User input:
0.000 to 999999 (6-digit floating-point number with sign)
!
Note!
The flow velocity must lie within the value range -12.5 m/s to +12.5 m/s.
Unit:
The unit configured in the measuring device is adopted.
Factory setting:
0.000
SIGNAL 1 PERIOD/
SIGNAL 2 PERIOD
Use this function to enter a duration for simulation signals 1 and 2 if you have selected
the STEP or PROFILE operating mode.
!
Note!
More information is provided in the OPERATION MODE function.
User input:
1 to 999 s (seconds)
Factory setting:
10 s
SIGNAL 1 PROFILE/
SIGNAL 2 PROFILE
For SIGNAL 1 PROFILE, enter the period of time for the change from simulation signal
1 to simulation signal 2. For SIGNAL 2 PROFILE, enter the period of time for the change
from simulation signal 2 to simulation signal 1.
!
Note!
• This function is only available if the PROFILE setting was selected in the OPERATION
MODE function.
• More information is provided in the OPERATION MODE function.
User input:
1 to 99 s (seconds)
Factory setting:
5 s
Function description
FUNCTION
SIMULATION
CONFIGURATION