Proline Fieldcheck
Wiring
Hauser
29
3.14
Connecting to Prosonic Flow 92 in twin-chamber
housing
"
Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the connection compartment.
1.
Remove the cover of the electronics compartment (6), remove the display module and lift up
the plastic cover.
2.
Disconnect the signal cable connectors (7) from the connection board.
3.
Connect the connectors of the Simubox cable to the adapter (3).
4.
Connect Fieldcheck to the current and frequency outputs of the transmitter (8) (terminal
pairs 22/23, 24/25 or 26/27).
!
Note!
The configuration of the outputs depends on the product type and the order version of the
tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
"Wiring" section of the flowmeter Operating Instructions.
5.
Connect Fieldcheck to the service connector (4).
A0013550
Fig. 22:
Connecting Fieldcheck to Prosonic Flow 92 (Ex d version)
1
Fieldcheck flow simulator
2
Simubox
3
Adapter, order number: 71110558
4
Service connector connection (use Fieldcheck/CDI connecting cable, order number 71032689
5
Connection compartment cover
6
Electronics compartment cover
7
Signal cable connector
8
Connecting cable for Fieldcheck/current and frequency outputs
ON
1 2 3 4 5 6 7 8
S
A
ON
1 2
4
0
1
2
3
4
5
6
7