Proline Fieldcheck
Wiring
Hauser
27
3.12
Connecting to Prosonic Flow 91 field housing
"
Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the connection compartment.
1.
Remove the cover of the connection compartment.
2.
Disconnect the sensor signal cables (4).
3.
Connect the connectors of the Simubox cable with the amplifier sockets and the Fieldcheck
socket (5).
!
Note!
The SYNC connector of the Simubox cable must be connected to the Fieldcheck socket.
4.
Connect Fieldcheck to terminal pair 24/25 for pulse and terminal pair 26/27 for power.
!
Note!
The configuration of the outputs depends on the product type and the order version of the
tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
"Wiring" section of the flowmeter Operating Instructions.
5.
Connect Fieldcheck to the service connector (6) using a special connecting cable.
Connecting cable order code: 71032689
a0009211
Fig. 20:
Connecting Fieldcheck to Prosonic Flow 91 (field housing)
1
Fieldcheck flow simulator
2
Simubox
3
Transmitter, field housing
4
Sensor cable
5
Amplifier sockets for sensor signal cable and Fieldcheck socket
6
Service connector (Fieldcheck/CDI connecting cable, use order number 71032689
3
5
6
2
1
4
Содержание Proline Fieldcheck
Страница 2: ......
Страница 103: ...Proline Fieldcheck Index Endress Hauser 103...
Страница 104: ...www endress com worldwide BA00067D 06 EN 13 11 71158188 FM SGML 9 0 ProMoDo...