Wiring
Proline Fieldcheck
18
Hauser
3.5.1
Connecting sensor test box to Promass 40, 80, 83, 84 field
housing
!
Note!
The start-up procedure must have been executed via the Simubox before connecting the sensor test
box.
"
Caution!
Connect the ground terminal of the adapter to the uninsulated housing components or to the
ground terminals in the connection compartment.
1.
Connect the sensor test box to Fieldcheck with the simulation cable.
2.
Connect the sensor test box to the sensor cable connector and the excitation coil connector
with the Coriolis adapter.
A0013401
Fig. 11:
Coriolis adapter for sensor test box
1
Sensor test box
2
Coriolis adapter
3
Excitation current connector Promass 40
4
Excitation current connector Promass 80, 83, 84
5
Signal coil connector Promass 40
6
Signal coil connector Promass 80, 83, 84
7
Ground terminal
8
Connection to Fieldcheck (with Simubox cable)
2
4
6
7
5
3
1
8