Micropilot S FMR533 4 to 20 mA HART
Commissioning
Hauser
23
5.3.1
Basic setup process
L00-FMR5xx-16-00-00-en-001
1)
Only selected tank geometrie "stilling well";
2)
Note: the dipped value should be taken at a level
≥
2 m (6.6 ft).
!
Note!
After finishing the basic setup pairs of values "measuring value Micropilot S - hand dipping value"
should be collected and if necessary a further correction should be performed through entering
characteristic pairs of values into the dip table. Information on the usage of the dip table can be
found in BA00209F/00/EN.
tank shape
medium property
process cond.
empty calibr.
full calibr.
pipe diameter
1)
dist. / meas. value
history reset
adjust “empty calibr.“
check distance
range of mapping
start mapping
According to
envelope curve
According to
hand dipping
2)
Distance
incorrect
Meas. value
incorrect
Distance
correct
Meas. value
correct