Commissioning
Micropilot S FMR533 4 to 20 mA HART
22
Hauser
"
Caution!
To successfully commission a precise measurement to the nearest mm, it is important you carry
out a
history reset
on
first installation
after mechanical installation and
after
the basic setup
of the device. Only after a history reset the
mounting calibration
is carried out. When a value
is dipped at a later date, make a second entry into the dip table, using the semi-automatic mode.
This way, you can easily carry out a linear correction of the measurement.
When configuring the function in
"basic setup"
(00) please take into account of the following
notes:
• Select the functions as descriped in BA209F/00/EN.
• Some functions can only be used depending on the parametrisation of the device. For
example, the pipe diameter of a stilling well can only be entered if
"stilling well"
was
selected beforehand in the
"tank shape"
(002) function.
• Certain functions (e.g. starting an interference echo mapping (053)) prompt you to confirm
your data entries. Press
O
or
S
to select
"YES"
and press
F
to confirm. The function is
now started.
• If you do not press a key during a configurable time period (
→
function group
"display"
(09)
), an automatic return is made to the home position (measured value display).
!
Note!
• The device continues to measure while data entry is in progress, i.e. the current measured
values are output via the signal outputs in the normal way.
• If the envelope curve mode is active on the display, the measured values are updated in a
slower cycle time. Thus, it is advisable to leave the envelope curve mode after the measuring
point has been optimised.
• If the power supply fails, all preset and parameterised values remain safely stored in the
EEPROM.
• All functions are described in detail, as is the overview of the operating menu itself, in the
manual
"BA00217F - Description of Instrument Functions"
, which can be found on the
enclosed CD-ROM.
• The default values of the parameters are typed in
boldface
.