Operation
Micropilot S FMR533 4 to 20 mA HART
16
Hauser
4
Operation
4.1
General structure of the operating menu
The operating menu is made up of two levels:
•
Function groups (00, 01, 03, …, 0C, 0D):
The individual operating options of the device are split up roughly into different function
groups. The function groups that are available include, e.g.:
"basic setup"
,
"safety settings"
,
"output"
,
"display"
, etc.
•
Functions (001, 002, 003, …, 0D8, 0D9):
Each function group consists of one or more functions. The functions perform the actual
operation or parameterisation of the device. Numerical values can be entered here and
parameters can be selected and saved. The available functions of the “
basic setup
"
(00)
function group include, e.g.:
"tank shape" (002)
,
"medium property" (003)
,
"process
cond." (004)
,
"empty calibr." (005)
, etc.
If, for example, the application of the device is to be changed, carry out the following procedure:
1.
Select the
"basic setup" (00)
function group
2.
Select the
"tank shape" (002)
function (where the existing tank shape is selected).
4.1.1
Identifying the functions
For simple orientation within the function menus, for each function a position is shown on the
display.
L00-FMRxxxxx-07-00-00-de-005
The first two digits identify the function group:
•
basic setup
00
•
safety settings
01
•
linearisation
04
...