Commissioning
Micropilot S FMR533 4 to 20 mA HART
28
Hauser
5.4.9
Function "dist./ meas. value" (008)
The
distance
measured from the reference point to the product surface and the
level
calculated
with the aid of the empty adjustment are displayed. Check whether the values correspond to the
actual level or the actual distance. The following cases can occur:
• Distance correct - level correct
È
continue with the next function
"check distance" (051)
• Distance correct - level incorrect
È
Check
"empty calibr" (005)
• Distance incorrect - level incorrect
È
continue with the next function
"check distance"
(051)
5.4.10 Function "check distance" (051)
This function triggers the mapping of interference echoes. To do so, the measured distance must
be compared with the actual distance to the product surface. The following options are available
for selection:
Selection:
• distance = ok
• dist. too small
• dist. too big
•
dist. unknown
• manual
⇒
⇒
E HAUSER
E
+
–
E HAUSER
E
+
–