Micropilot S FMR533 4 to 20 mA HART
Commissioning
Hauser
31
5.4.14 Function "history reset" (009)
By this function a history reset of the device is performed, i.e. the correspondance table between
level and index values is deleted. A new correspondance table will be filled and stored after the
history reset.
"
Caution!
A history reset must be performed after:
• first installation or
• change of basic setup or
• change of the installation situation.
In this case also perform a reset of the dip table in function
"dip table mode" (033)
.
!
Note!
After the basic setup, an evaluation of the measurement with the aid of the envelope curve
(
"envelope curve" (0E)
function group) is recommended.
⇒