Code
Type
Cause and Remedy
E101
…
E106
Alarm
Fault in instrument electronics
- Check supply voltage
- Call Hauser Service
E201
Alarm
Fault in detector, pulse rate too low
(DG 17 = 0, DG 27 < 1 Hz, DG 57 < 2 Hz)
- Check detector
E202
Alarm
Pulse rate > max. possible pulse rate (detector possibly defective)
(DG 17/27 > 2 x empty calibration countrate; DG 57 ca. 110% of
max. countrate - dependent on temperature and calibration value)
- If calibration was made with build-up on the sides of the vessel, it is
possible that this has fallen off and less radiation is being absorbed.
In this case recalibrate and set V0H1 equal to e.g. 20.
- Check detector
- Check for external radiation source
E204
Alarm
No temperature measurement (DG 57 only)
E206
Alarm
LED reference measurement outside limits 30
…
95% (DG 57 only)
E401
Alarm
No signal from detector,
- Check the wiring. Take into account the polarity.
- If the wiring is correct, check the voltage at the terminals of the
detector. The value should be 13.7 +/- 3V.
If the voltage is correct, a mistake is present at the detector. Exchange
detector.
- If the voltage is not correct at the detector, check for the same
voltage at the output of the transmitter FMG 671. If the output voltage
is not correct, the transmitter is defective.
E404
Alarm
Wrong detector connected
E610
Warning
Detector data have been changed (DG 57 only)
- Check whether detector data chip (DAT) is inserted in card
or whether it corresponds to the detector connected
-- Re-calibrate, if detector exchanged
E613
Warning
Instrument in simulation mode
- Switch back when finished
E615
Alarm
Incorrect calibration points, e.g. through mix-up, repeat calibration
E630
Warning
Calibration incomplete
E631
Warning
Limit for switch point relay 1 exceeds maximum or drops below
minimum countrate
- re-enter
E632
Warning
Limit for switch point relay 2 exceeds maximum or drops below
minimum countrate
- re-enter
E633
Warning
Hysteresis relay 1 too small
- re-enter, increase integration time
E634
Warning
Hysteresis relay 2 too small
re-enter, increase integration time
E635
Warning
Difference between maximum and mimimum pulse rate too small
- increase integration time
- source too weak
E660
Warning
Detector temperature > 55°C (DG 57 only)
Table 7.1
Error codes and error messages
Gammapilot FTG 671
Chapter 7: Trouble-Shooting
50
Содержание gammapilot FTG 671
Страница 4: ...Gammapilot FTG 671 Table of Contents 2...
Страница 12: ...Gammapilot FTG 671 Chapter 1 Introduction 10...
Страница 28: ...Gammapilot FTG 671 Chapter 2 Installation 26...
Страница 32: ...Gammapilot FTG 671 Chapter 3 Controls 30...
Страница 48: ...Gammapilot FTG 671 Chapter 5 Limit Switches Relays 46...
Страница 50: ...Gammapilot FTG 671 Chapter 6 Maintenance 48...
Страница 59: ...Gammapilot FTG 671 Operating Matrix 57...
Страница 61: ...Index Gammapilot FTG 671 59...
Страница 62: ...Gammapilot FTG 671 Index 60...
Страница 63: ...Index Gammapilot FTG 671 61...