Detector DG 27 - Co 60
4 HVL
max. attenuation
response time (s)
local
d
o
se rate at d
e
tector
[
µ
Sv/h]
Fig. 5.12:
Relay response time as a
function of local dose rate for
detector DG 27 and source Co 60
Detector DG 27 - Cs 137
4 HVL
max. attenuation
local
dose rat
e
at
detect
or
response time (s)
BA082D67
[
µ
Sv/h]
Fig. 5.13:
Relay response time as a
function of local dose rate for
detector DG 27 and source
Cs 137
0
5
10
15
20
25
10
20
re
sponse t
im
e
(s)
dose rate factor
BA082D67
0.5 HVL
0.7 HVL
1 HVL
2 HVL
4 HVL
Cs 137
Factor for DG 57/100: see graph
Factor for DG 57/400: graph x 0.5
Co 60
Factor for DG 57/100: graph x 1.8
Factor for DG 57/400: graph x 0.9
1
Fig. 5.14:
Local dose rate as a function of
relay response time for
DG 57/100 and DG 57/400
detectors
Gammapilot FTG 671
Chapter 5: Limit Switches - Relays
44