Rackbus
For system use, wire the Rackbus as shown in Fig. 2.15.
•
For ZA 672 (Modbus) - Computer/PLC wiring see Operating Instructions
BA 054E
for ZA 673 (PROFIBUS) see Operating Instructions BA 073E
for ZA 674 (FIP) see Operating Instructions BA 085E
•
Be sure that the system is properly grounded. A difference in ground potential
between the network device and Gateway can cause loss of data or
transmission faults
Relays
Alarms and limit switch relays can be connected as shown in Fig. 2.16.
•
Max. current per contact 2.5A : for switching capacity see Section 2.6.
Alternative 2
+
RACKBUS
d b z
2
4
6
10 +
8
12 + +
14
16
32 L+
+
RACKBUS
d b z
2
4
6
10 +
8
12 + +
14
16
32 L+
+
RACKBUS
d b z
10 +
18 r
20 u a
22 r
24 u a
26 r
28 u a
30 L-
18 r
20 u a
22 r
24 u a
26 r
28 u a
30 L-
L+
L-
2
4
6
8
14
16
18
20
22
28
26
24
30
32
12
u
a
r
u
a
r
u
a
r
BA082E17
Max. 64
Commutec-
transmitters in
parallel on
Rackbus
Commutec transmitters FTG, FMB, FMC, FMU, PMX, TMT etc.
ZA 672
FTG 671
Fig. 2.15:
Rackbus wiring diagram
Max.current
per line
2,5 A
BA0082E18
Recommended
Alternative 1
Fig. 2.16:
Suggestions for wiring
together relays and alarms
Gammapilot FTG 671
Chapter 2: Installation
22
Содержание gammapilot FTG 671
Страница 4: ...Gammapilot FTG 671 Table of Contents 2...
Страница 12: ...Gammapilot FTG 671 Chapter 1 Introduction 10...
Страница 28: ...Gammapilot FTG 671 Chapter 2 Installation 26...
Страница 32: ...Gammapilot FTG 671 Chapter 3 Controls 30...
Страница 48: ...Gammapilot FTG 671 Chapter 5 Limit Switches Relays 46...
Страница 50: ...Gammapilot FTG 671 Chapter 6 Maintenance 48...
Страница 59: ...Gammapilot FTG 671 Operating Matrix 57...
Страница 61: ...Index Gammapilot FTG 671 59...
Страница 62: ...Gammapilot FTG 671 Index 60...
Страница 63: ...Index Gammapilot FTG 671 61...