7.3
Simulated operating mode
This function is intended primarily for checking the correct function of the system and is
selected and terminated at V8H0:
•
Enter 3 to simulate the input variable (pulse rate or frequency)
- display 0
…
9999, actual 0
…
99990
•
Enter 4 to simulate the normalised input variable (0
…
100)
•
Enter 0 to terminate simulation and resume normal measurements.
Start and stop simulation
Step
Matrix
Entry
Significance
1
V8H0
e.g.3
Selects simulation of input variable
or
e.g. 0
Selects standard operation mode and ends simulation
2
-
»E«
Registers entry
When a value is entered at matrix position V9H6 or V9H7, the relay response can be
tested accordingly. Throughout the simulation the red alarm LED flashes to indicate that
the instrument is no longer measuring, the alarm relay does not, however, trip.
Countrate/frequency
The countrate (or frequency) is entered in the same manner as it is displayed at V0H8,
i.e. as value/10, e.g.
•
120.1 = 1201 pulses/s, 1234 = 12340 pulses/s for DG 57
•
2.5 = 25 Hz, 0.9 = 9 Hz for DG 17/DG 27
Step
Matrix
Entry
Significance
1
V9H6
e.g. 100 1000 pulse/sec for DG 57 or 1000 Hz for DG 17/27
2
-
»E«
Registers entry
Normalised countrate
Any value between 0
…
100 can be entered
Step
Matrix
Entry
Significance
1
V9H7
e.g. 10
Normalised countrate/frequency = 10
2
-
»E«
Registers entry
Gammapilot FTG 671
Chapter 7: Trouble-Shooting
52
Содержание gammapilot FTG 671
Страница 4: ...Gammapilot FTG 671 Table of Contents 2...
Страница 12: ...Gammapilot FTG 671 Chapter 1 Introduction 10...
Страница 28: ...Gammapilot FTG 671 Chapter 2 Installation 26...
Страница 32: ...Gammapilot FTG 671 Chapter 3 Controls 30...
Страница 48: ...Gammapilot FTG 671 Chapter 5 Limit Switches Relays 46...
Страница 50: ...Gammapilot FTG 671 Chapter 6 Maintenance 48...
Страница 59: ...Gammapilot FTG 671 Operating Matrix 57...
Страница 61: ...Index Gammapilot FTG 671 59...
Страница 62: ...Gammapilot FTG 671 Index 60...
Страница 63: ...Index Gammapilot FTG 671 61...