4.6
Additional functions
Detector monitor
The Gammapilot FTG 671 also checks whether the detector is operating correctly.
Depending upon the type used, the following alarms and warnings may appear.
•
E 201:
Pulserate too low
Alarm
•
E 202:
Pulserate too high
Alarm
•
E 204:
Temperature measurement faulty (DG 57 only)
Alarm
•
E 206:
Reference measurement faultly (DG 57 only)
Alarm
•
E 401:
No signal from detector
Alarm
•
E 404:
Wrong detector connected
Alarm
•
E 610:
Detector data chip changed (DG 57 only)
Warning
- e.g. after exchange of detector
•
E 660:
Detector temperature too high (DG 57 only)
Warning
The self-monitoring function is described in detail in Chapter 6, »Trouble-Shooting«
4.7
Measured value display
During normal operation the normalised countrate can be read at V0H0. In addition to
this, several other fields contain system information which might be needed, e.g., for
trouble-shooting etc.. Table 4.3 summarizes the measured value displays.
Matrix
Measured value
Remarks
V0H0
Normalised pulse rate or
frequency
Display 0
…
100 of normalised countrate(PCM) or
frequency (PFM) depending upon the detector mode
used
V0H8
Current measuring
frequency or pulse rate
Displays the frequency or pulse rate which is actually
measured. The value is dispayed as frequency/10 or
counts/100ms
V0H9
Background radiation
DG 57 only
Pulse rate of the background radiation for calibration
mode 5. For calibration modes 1
…
4 is always 0.
Displayed as counts/100ms
V3H6
Free value
Original pulse rate or frequency used for calibration
V3H7
Covered value
Original pulse rate or frequency used for calibration
V3H8
Dose rate
Current mean dose rate measured at detector in
µ
Sv/h
V7H1
…
V7H9
Service parameters
Parameters which may be required for fault
diagnosis during a service call
V9H0
Current error code
When red fault LED flashes, the error code can be
read here
V9H1
Last error code
The previous error can be read and deleted here -
press »E«
V9H3
Software version with
instrument code
The first two figures indicate the instrument, the last,
the software version; e.g. DDXX:
Instrument code DD, Software version X.X
V9H4
Rackbus address
Indicates address set at DIP-switches on card
Table 4.9:
Matrix positions of measured
value displays
Gammapilot FTG 671
Chapter 4: Calibration and Operation
37
Содержание gammapilot FTG 671
Страница 4: ...Gammapilot FTG 671 Table of Contents 2...
Страница 12: ...Gammapilot FTG 671 Chapter 1 Introduction 10...
Страница 28: ...Gammapilot FTG 671 Chapter 2 Installation 26...
Страница 32: ...Gammapilot FTG 671 Chapter 3 Controls 30...
Страница 48: ...Gammapilot FTG 671 Chapter 5 Limit Switches Relays 46...
Страница 50: ...Gammapilot FTG 671 Chapter 6 Maintenance 48...
Страница 59: ...Gammapilot FTG 671 Operating Matrix 57...
Страница 61: ...Index Gammapilot FTG 671 59...
Страница 62: ...Gammapilot FTG 671 Index 60...
Страница 63: ...Index Gammapilot FTG 671 61...