4
Calibration and Operation
This chapter is concerned with the basic settings of the Gammapilot, the principle
sections describing the:
•
Commissioning
•
Standard calibration with background calibration
•
Standard calibration without background calibration
•
Provisional calibration
•
Limit detection
•
Additional functions
•
Display of measured values
•
Locking the parameter matrix.
The Gammapilot FTG 671 controls are described in Chapter 3, p. 23, and the setting of
the relays in Chapter 5, p. 35.
Note your settings!
Note your parameters in the tables provided in the back cover. Should the transmitter be
exchanged at a later date, simply re-enter all the parameters and the system is ready to
run - see Chapter 6, p. 42.
4.1
Commissioning
Note!
•
Make sure that the instrument has been switched on for at least six hours before it is
commissioned or calibrated.
The commissioning comprises two steps:
•
Transmitter reset
•
Entry of detector information prior to calibration
Transmitter reset
A transmitter reset enters the factory based parameters into the operating matrix, see
Table in back cover. It is made by entering a number between 670 and 679 at V9H5.
Step
Matrix
Entry
Significance
1
V9H5
e.g. 671
Enter any number 670
…
679 to reset transmitter
2
-
»E«
Registers entry
Gammapilot FTG 671
Chapter 4: Calibration and Operation
31
Содержание gammapilot FTG 671
Страница 4: ...Gammapilot FTG 671 Table of Contents 2...
Страница 12: ...Gammapilot FTG 671 Chapter 1 Introduction 10...
Страница 28: ...Gammapilot FTG 671 Chapter 2 Installation 26...
Страница 32: ...Gammapilot FTG 671 Chapter 3 Controls 30...
Страница 48: ...Gammapilot FTG 671 Chapter 5 Limit Switches Relays 46...
Страница 50: ...Gammapilot FTG 671 Chapter 6 Maintenance 48...
Страница 59: ...Gammapilot FTG 671 Operating Matrix 57...
Страница 61: ...Index Gammapilot FTG 671 59...
Страница 62: ...Gammapilot FTG 671 Index 60...
Страница 63: ...Index Gammapilot FTG 671 61...