Deltabar PMD75B HART
Technical data
Hauser
73
Vibration resistance
Single compartment housing
Measuring range
Sine wave oscillation IEC 61298-3:2008 Shock
10 mbar (0.15 psi) and 30 mbar (0.45 psi) (only up to
PN100)
10 Hz to 60 Hz: ±0.21 mm (0.0083 in)
60 Hz to 2000 Hz: 3 g
30 g
0.1 to 250 bar (1.5 to 3
750 psi)
10 Hz to 60 Hz: ±0.35 mm (0.0138 in)
60 Hz to 1000 Hz: 5 g
30 g
Aluminum dual compartment housing
Measuring range
Sine wave oscillation IEC 61298-3:2008
Shock
10 mbar (0.15 psi) and 30 mbar (0.45 psi)
10 Hz to 60 Hz: ±0.21 mm (0.0083 in)
60 Hz to 2000 Hz: 3 g
30 g
0.1 to 250 bar (1.5 to 3
750 psi)
10 Hz to 60 Hz: ±0.35 mm (0.0138 in)
60 Hz to 1000 Hz: 5 g
30 g
Stainless steel dual compartment housing
Measuring range
Sine wave oscillation IEC 61298-3:2008 Shock
10 mbar (0.15 psi) and 30 mbar (0.45 psi) (only up to
PN63)
10 Hz to 60 Hz: ±0.15 mm (0.0059 in)
60 Hz to 500 Hz: 1 g
15 g
0.1 to 250 bar (1.5 to 3
750 psi)
10 Hz to 60 Hz: ±0.15 mm (0.0059 in)
60 Hz to 500 Hz: 2 g
15 g
Dual compartment housing, L-shape
Sine wave oscillation IEC 61298-3:2008
Shock
10 Hz to 60 Hz: ±0.21 mm (0.0083 in)
60 Hz to 2000 Hz: 3 g
30 g
Electromagnetic
compatibility (EMC)
• Electromagnetic compatibility as per EN 61326 series and NAMUR recommendation
EMC (NE21)
• With regard to the safety function (SIL), the requirements of EN 61326-3-x are satisfied
• Maximum deviation with interference influence: < 0.5% of span with full measuring
range (TD 1:1)
For more details refer to the EU Declaration of Conformity.