Commissioning
Deltabar PMD75B HART
40
Hauser
9.7.2
Commissioning with an operating menu
All configuration tools offer a
Commissioning
wizard that helps the user to configure
the most important configuration parameters (
Guidance
menu
Commissioning
wizard).
Commissioning comprises the following steps:
• Position adjustment/zero adjustment
• Configuring measurement
Performing position adjustment/zero adjustment with the local display
Perform position adjustment/zero adjustment if necessary:
• In the main menu, select the
Guidance
menu
Commissioning
wizard with the or
key
Press the key
• Select the
Zero adjustment
parameter with the key
Press the key until you
Zero adjustment
parameter is reached
Press the key
• Select the
Confirm
option with the key
• Press the key
Position adjustment/zero adjustment has been performed.
Press the key until the end of the Wizard is reached and press the key to exit the
Wizard.
Back to the main menu: press the and key simultaneously (2 x).
Performing position adjustment/zero adjustment with FieldCare
Perform position adjustment/zero adjustment if necessary.
In the
Zero adjustment
parameter, select the
Confirm
option and press Enter.
In the
Application
menu, select the
Sensor
submenu
Sensor calibration
submenu
Zero
adjustment
parameter and press Enter.
Position adjustment/zero adjustment has been performed.