10.1 Reading the device locking status
Display of active write protection:
Locking status
parameter
Menu path of local display: at the top operating level
Menu path of operating tool: System → Device management
10.2 Reading measured values
With the
Measured values
submenu, it is possible to read all the measured values.
Navigation
"Application" submenu → Measured values
10.3 Adapting the measuring device to the process
conditions
The following are available for this purpose:
• Basic settings using the
Guidance
menu
• Advanced settings using the
Diagnostics
menu,
Application
menu and
System
menu
10.3.1 Sensor calibration
In the course of their life cycle, pressure sensors
can
deviate, or drift,
from the original
pressure characteristic curve. This deviation is caused by physical factors and can be
corrected with a sensor calibration.
Absolute pressure sensors cannot be recalibrated.
Perform the sensor calibration. Adhere to the sequence indicated!
1. Apply the lower pressure value to the device.
2. Enter the value for the
Lower sensor trim
parameter. Application → Sensor → Sensor
calibration → Lower sensor trim
The value entered causes a parallel shift of the pressure characteristic in relation
to the current calibration.
3. Apply the upper pressure value to the device.
4. Enter the value for the
Upper sensor trim
parameter. Application → Sensor → Sensor
calibration → Upper sensor trim
The value entered causes a change in the slope of the current calibration curve.
2)
Not possible with display.
3)
Deviations caused by physical factors are also known as "Sensor drift".