Repair
Deltabar PMD75B HART
66
Hauser
13
Repair
13.1 General information
13.1.1 Repair concept
Under the Hauser repair concept, devices have a modular design and repairs are
carried out by Hauser Service or by properly trained customers.
Spare parts are grouped into logical kits with the associated replacement instructions.
For more information on service and spare parts contact Hauser Service.
13.1.2 Repair of Ex-certified devices
L
WARNING
Limitation of electrical safety due to incorrect repair!
Risk of explosion!
‣
Only specialist personnel or Hauser-Service can carry out repairs to Ex
certified devices.
‣
Relevant standards and national regulations as well as safety instructions (XA) and
certificates must be observed.
‣
Only genuine Hauser spare parts may be used.
‣
When ordering spare parts, please check the device designation on the nameplate. Only
identical parts may be used as replacements.
‣
Carry out repairs according to the instructions. Following a repair, the device must
fulfill the requirements of the individual tests specified for that device.
‣
A certified device may be converted to another certified device version by Endress
+Hauser Service only.
‣
All repairs and modifications must be documented.
13.2 Spare parts
• Some replaceable measuring device components are identified by means of a spare part
nameplate. This contains information about the spare part.
• All the spare parts for the measuring device, along with the order code, are listed in the
W@M Device Viewer
(
) and can be ordered. If available,
users can also download the associated Installation Instructions.
Measuring device serial number:
• Located on the device and spare part nameplate.
• Can be read out via the device software.