A large standardization offset may be caused by a poisoned reference electrode. Poisoning
agents can cause the pH to be offset by as much as two pH units. To check the reference
voltage, see
.
AdC
AdC means the analog to digital converter has failed.
1.
Verify that the sensor wiring is correct and connections are tight. Be sure to check
connections at the junction box if one is being used.
2.
Disconnect sensor(s) and simulate temperature and sensor input.
To simulate
See Section
Dissolved oxy-
gen
Ozone or chlor-
ine
pH
Temperature
If the transmitter does not respond to simulated signals, the analog PCB has
probably failed.
3.
Call the factory for assistance.
bAd Gnd
bAd Gnd usually means a problem with the analog PCB. Contact the factory.
In too biG
In too biG means the raw millivolt signal from the pH sensor is too large.
1.
Verify that sensor wiring is correct and connections are tight. Be sure to check
connections at the junction box if one is being used.
2.
Replace the pH sensor with a sensor known to be working.
3.
If replacing the pH sensor does not cause the message to disappear, call the factory
for assistance.
RitE Err
Program settings in the 5081-A can be protected against accidental changes by setting a
three-digit security code. Settings can be further protected by removing a jumper (JP-1)
from the CPU board. If JP-1 has been removed, program settings cannot be changed.
Diagnostics and troubleshooting
98
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...