if the transmitter responds to simulated inputs, the problem must lie with the
sensor or the interconnecting wiring. Verify the interconnecting wiring point to
point. Fix or replace bad cable. If cable is good, replace the pH sensor.
Std Err
Std Err
means the reference electrode voltage has changed drastically. Typical causes are
exposure to poisoning agents, sulfides or cyanides, or prolonged exposure to high
temperature.
Flowchart-Std Err
Troubleshooting depends on the type of sensor.
•
If the sensor is rebuildable, replenish the electrolyte solution and replace the liquid
junction. Calibrate the sensor.
-
If the sensor can be calibrated, the problem has been corrected.
-
If the sensor cannot be calibrated, replace the sensor. If the sensor has separate
measuring and reference electrodes, replace only the reference electrode.
•
If the sensor is not rebuildable, try the reference electrode rejuvenation procedure.
-
If the rejuvenated sensor can be calibrated, the problem has been corrected.
-
If the sensor cannot be calibrated, replace the sensor.
rOM FAIL or CPU FAIL
rOM FAIL or CPU FAIL means the transmitter electronics have failed. Replace the electronic
board stack (PN 23992-02 [-HT] or PN 23992-03 [-HF]).
AdC WArn or CyCLE PWr
The Adc WArn or CyCLE PWr message appears momentarily when the transmitter has
recognized an internal calculation problem. The transmitter repeats the calculation, and
the message disappears once the calculation is successful. If the message is displayed
constantly, the transmitter electronics may be faulty.
Troubleshooting AdCWArn or CyCLE PWr
1.
Check transmitter performance by simulating pH inputs. See
•
If the transmitter performs satisfactorily, go to Step 2.
•
If the transmitter does not respond to simulated inputs, replace the board stack.
2.
If the transmitter responds to simulated inputs, the problem must lie with the
sensor or the interconnecting wiring. Verify the interconnecting wiring point to
point. Fix or replace bad cable. If cable is good, replace the pH sensor.
WritE Err
WriTE Err
means that jumper JP1 on the CPU board is not in place. If the sensor is not in
place, the transmitter cannot be programmed or calibrated.
Diagnostics and troubleshooting
Instruction Manual
67
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...