Chlorine readings too low
1.
Was the sample tested as soon as it was taken? Chlorine solutions are unstable. Test
the sample immediately after collecting it. Avoid exposing the sample to sunlight.
2.
Low readings can be caused by zeroing the sensor before the residual current has
reached a stable minimum value. Residual current is the current the sensor
generates even when no chlorine is in the sample. Because the residual current is
subtracted from subsequent measured currents, zeroing before the current is a
minimum can lead to low results. See
3.
Sensor response depends on flow. If the flow is too low, readings will be low and flow
sensitive. Verify that the flow past the sensor equals or exceeds the minimum value.
See the sensor instruction manual for recommended flows.Verify that the chlorine
sensor is installed in the correct flow cell. See Figure and Figure. Verify that the liquid
level in the constant head sampler is level with the central overflow tube and that
excess sample is flowing down the tube. If necessary, disassemble and clean the
overflow sampler. See
Section
.
8.4.6
Total chlorine measurement and calibration problems
Refer to the instruction manual for the SCS921 for a complete troubleshooting guide.
8.4.7
Ozone measurement and calibration problems
Problem
See Section
Zero current is substantially outside the
range -10 to 10 nA.
Ozone zero current outside range
Zero reading is unstable.
Sensor current during calibration is sub-
stantially less than about 350 nA/ppm at
25 °C (77 °F).
Sensor current less than 350 nA/ppm
Process readings are erratic.
Ozone process readings erratic
Readings drift.
Sensor does not respond to changes in
ozone level.
Sensor not responding to ozone changes
Ozone readings are too low.
Ozone zero current outside range
The ozone zero current is substantially outside the range -10 to 10 nA.
1.
Is the sensor properly wired to the transmitter?
2.
Is the zero solution ozone free? Test the zero solution for ozone level. The
concentration should be less than 0.02 ppm.
Diagnostics and troubleshooting
106
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...