1.
Place the sensor in a solution of 5% sodium sulfite (Na
2
SO
3
) in water. Ensure that air
bubbles are not trapped against the membrane.
2.
Go to the main display; press
DIAG
and then
NEXT
.
The SENSOR CUR prompt appears.
3.
Press
ENTER
to view the sensor current. Make sure the sensor reaches its zero
current.
This may require several hours. Do not start the zero routine until the sensor has
been in zero solution for at least two hours.
4.
Press
CAL
on the IRC.
The SENSOR 0 prompt appears.
5.
Press
ENTER
.
The screen shows the appropriate value correlating to the zero value. The screen
shows 0.02. The reading must be below 0.02 ppm for the zero calibration to be
accepted.
6.
To change the zero limit, use the editing keys and then press
ENTER
.
The TIME DELAY message appears and remains until the zero current is below the
concentration limit. If the current is already below the limit, TIME DELAY will not
appear.
7.
To bypass the time delay, press
ENTER
.
When the procedure is complete, 0 DONE appears.
8.
Press
EXIT
.
9.
Press
RESET
to return to the main display.
7.4.3
Air calibration procedure (dissolved oxygen)
1.
Remove the sensor from the process liquid. Use a soft tissue and a wash bottle to
clean the membrane. Dry it by blotting.
The membrane must be completely dry during air calibration.
2.
Pour some water into a beaker and suspend the sensor with the membrane
approximately 1/2 in. (approximately 1 cm) above the water surface.
Keep the sensor out of direct sunlight.
3.
Monitor the dissolved oxygen and temperature reading. Once the readings are
relatively stable, begin the calibration.
It may take 10 - 15 minutes for the sensor reading in the air to stabilize.
4.
Press
CAL
on the IRC.
5.
Press
NEXT
.
Measurements
Instruction Manual
49
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...