c. Connect a jumper wire between TB-3 (RTD RTN) and TB-4 (RTD-SNS). Connect a
second jumper wire between TB-7 (REF IN) and TB-8 (SOL GND).
d. Place the sensor back in the process liquid. If diagnostic measures such as
GLASSFAIL or REF WArn appear, turn off the sensor diagnostics.
•
If the symptoms disappear, interference was coming into the transmitter
along one of the sensor wires. The measurement system can be operated
permanently with simplified wiring.
•
If symptoms still persist, go to Step 4.
4.
Check for extra ground connections or reduced noise.
a. The electrode system is connected to earth ground through the process. If other
ground connections exist, there are multiple paths and ground loops are present.
Noise enters the measurement either by a direct connection, usually between
the cable and grounded metal, or by an indirect connection, usually EMI/RFI
picked up by the cable.
b. If the sensor cable is run inside conduit, there may be a short between the cable
and the conduit. Re-run the cable outside the conduit. If symptoms disappear,
then a short exists between the table and the conduit. Likely a shield is exposed
and is touching the conduit. Repair the cable and reinstall it in the conduit.
c. To avoid induced noise in the sensor cable, run it as far away as possible from
power cables, relays, and electric motors. Keep sensor wiring out of crowded
panels and cable trays.
d. Occasionally, noise can travel into the transmitter housing from the metal it is
mounted on. The noise is then radiated into the transmitter electronics. If
isolating the transmitter from its metal mounting eliminates the symptoms,
move the transmitter to a different location or mount it with isolating materials.
e. If ground loop problems persist, consult the factory. A visit from an experienced
service technician may be required to solve plant-induced problems.
Temperature reading is inaccurate
1.
To troubleshoot temperature problems, refer to
.
2.
Calibrate the temperature response of the sensor.
3.
If necessary, automatic temperature compensation can be temporarily disabled and
the transmitter placed in manual temperature compensation. For manual
temperature, choose a temperature equal to the average temperature of the
process. The resulting pH reading will be in error. The more variable the
temperature and the further from pH 7, the greater the error.
HART communications problems
1.
If the 375 or 475 communicator software does not recognize the 5081 pH/ORP
transmitter, order an upgrade from Rosemount at 800 999 9307.
2.
Be sure the HART load and voltage requirements are met.
a. HART communications requires a minimum 250 ohm load in the current loop.
b. Install a 250 - 500 ohm resistor in series with the current loop. Check the actual
resistor value with an ohmmeter.
Diagnostics and troubleshooting
72
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...