8
Diagnostics and troubleshooting
8.1
pH/ORP diagnostics and troubleshooting
8.1.1
5081-P warning and fault messages
The 5081-P pH/ORP transmitter continuously monitors the measurement loop (sensor and
transmitter) for conditions that cause erroneous measurements. When a problem occurs,
the transmitter displays either a warning or fault message. A warning alerts you that a
potentially system disabling condition exists. If the condition causing the problem is not
corrected, there is a high probability that the system will soon fail. A fault alerts you that a
system disabling condition exists. If a fault message is showing, all measurements should
be regarded as erroneous.
When a Warning condition exists:
1.
The main display remains stable; it does not flash.
2.
A warning message appears alternately with the temperature display. See
.
for an explanation of the different warnings and suggested ways of
correcting the problem.
Warning annunciation
Figure 8-1:
When a Fault exists:
1.
The main display flashes.
2.
The words
FAULT
and
HOLD
appear in the main display.
3.
A fault message appears alternately with the temperature/output display. See
for an explanation of the different fault messages and
suggested ways of correcting the problem.
4.
The output current will remain at the present value or go to the programmed fault
value.
Diagnostics and troubleshooting
Instruction Manual
53
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...