5.
If the transmitter is in Hold when the fault occurs, the output remains at the
programmed hold value. To alert you that a fault exists, the word
FAULT
appears in
the main display, and the display flashes. A fault or diagnostic message also appears.
6.
If the transmitter is simulating an output current when the fault occurs, the
transmitter continues to generate the simulated current. To alert you that a fault
exists, the word
FAULT
appears in the display, and the display flashes.
Fault annunciation
Figure 8-2:
When a disabling condition, a fault, occurs, the display appears as pictured above. To further alert you
that measurements are in error, the display flashes. Diagnostic messages appear in the temperature/
output area on the screen.
8.1.2
Calibration errors
If an error occurs during calibration, an error message appears in the main display, and the
transmitter does not update the calibration. The calibration errors are
Std Err
,
SLOPE Err LO
,
and
SLOPE Err HI
. See
for an explanation of the error messages and suggested
ways of correcting the problem.
8.1.3
Troubleshooting - general
Follow the steps below to troubleshoot your transmitter.
1.
Look for a diagnostic message on the display to help identify the problem.
Refer to
for an explanation of the message and a list of the possible
problems that triggered it.
2.
Refer to
for common measurement problems and the recommended
actions to resolve them.
3.
Follwo the step-by-step troubleshooting approach, offered in
diagnose and correct less common or more complex problems.
Diagnostics and troubleshooting
54
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...