3
Installation
3.1
Unpacking and inspection
Complete the following steps when you unpack your instrument.
1.
Inspect the shipping container. If there is damage, contact the shipper immediately
for instructions.
2.
If there is no apparent damage, remove the transmitter.
3.
Ensure that all items shown on the packing list are present. If items are missing,
contact your local Customer Care representative.
4.
Save the shipping container and packaging.
They can be reused to return the transmitter to the factory in case of damage.
3.2
Installation
3.2.1
General information
1.
The transmitter tolerates harsh environments. For best results, install the
transmitter in an area where temperature extremes, vibrations, and
electromagnetic and radio frequency interference are minimized or absent.
2.
To prevent unintentional exposure of the transmitter circuitry to the plant
environment, keep the security lock in place over the circuit end cap. To remove the
circuit end cap, loosen the lock nut until the tab disengages from the cap end and
then unscrew the cover.
3.
The transmitter has two 3/4 inch conduit openings, one on each side of the housing.
Run sensor cable through the left side opening (as viewed from the wiring terminal
end of the transmitter) and run power wiring through the right side opening.
4.
Use water tight cable glands to keep moisture out of the transmitter.
5.
If conduit is used, plug and seal the connections at the transmitter housing to
prevent moisture from getting inside the transmitter.
CAUTION!
EQUIPMENT DAMAGE
Moisture accumulating in the transmitter housing can affect the performance of the
transmitter and may void the warranty.
6.
If the transmitter is installed some distance from the sensor, a remote junction box
with preamplifier in the junction box or in the sensor may be necessary. Consult the
sensor instruction manual for maximum cable lengths.
Installation
Instruction Manual
11
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...