The SENSOR CAL submenu appears.
6.
Press
ENTER
.
The AIR CAL prompt appears.
7.
Press
ENTER
.
8.
Select the units. Then press
NEXT
.
9.
Use the arrow keys to enter the barometric pressure. Press
ENTER
when done.
The TIME DELAY message appears and remains until the sensor reading is relatively
stable.
10.
To bypass the time delay, press
ENTER
.
CAL DONE appears when the calibration is complete.
11.
Press
EXIT
.
12.
To return to the main display, press
RESET
.
7.4.4
In-process calibration procedure (dissolved oxygen)
The sensor can be calibrated against a standard instrument.
1.
Press
CAL
on the IRC.
2.
Press
NEXT
.
The SENSOR CAL submenu appears.
3.
Press
ENTER
.
4.
Press
NEXT
.
The AIR CAL prompt appears.
5.
Press
NEXT
.
The INPROCESS prompt appears.
6.
Press
ENTER
.
The TIME DELAY message appears and will remain until the sensor is relatively
stable. To bypass the time delay, press
ENTER
. The GRAP SPL message appears.
7.
Press
ENTER
.
8.
Use the arrow keys to change the flashing display to the value indicated by the
standards instrument. Press
ENTER
.
9.
Press
RESET
to return to the main display.
Measurements
50
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...