NOTICE
GLASSFAIL
is a sensor diagnostic message. Sensor diagnostic messages are optional. They can
be turned off.
Troubleshooting flowchart - GLASSFAIL
1.
Be sure the sensor is completely immersed in the process liquid.
If the diagnostic message disappears, the sensor is in good condition.
If the diagnostic message remains, go to Step 2.
2.
Measure the glass impedance. See
for the procedure. Note the reading.
•
If the glass impedance is low (< 40 megohms)...
a. Check preamp location in program menus (PAMP = _____).
-
If the location is incorrect, go to Step b.
-
If after selecting the correct location of Preamp in the Program menu, the
glass impedance is still low, go to Step b.
b. Calibrate the sensor.
-
If the sensor calibrates properly...
a. The sensor is in good condition, but the Glass Fail Low (GFL) limit is set
too high.
b. Lower the GFL limit to about 10 megohms below the glass impedance
value (GIMP) measured in Step 2.
c. If the Glass Warning Low (GWL) message was also flashing, lower the
limit from its former value by the same amount the GFL was lowered
from its former value.
-
If the sensor cannot be calibrated, the pH glass membrane is likely cracked
and the sensor must be replaced. The crack in the glass may not be visible
or may be difficult to see.
•
If the glass impedance is high (> 800 megohms)...
a. Check that the sensor is correctly wired to the transmitter. Pay particular
attention to the following:
a. For Rosemount PLUS (+) and TUpH sensors with integral preamplifiers, the
blue solution ground wire must be attached to TB-8 (SOL GND), and the
gray reference in wire must be attached to TB-7 (REF IN).
NOTICE
TB-8 means terminal 8 on the terminal board.
b. If the sensor was wired with the blue solution ground wire unattached and
a jumper between the terminals TB-8 and TB-7, remove the jumper and
reattach the blue solution ground wire to TB-8. Keep the gray reference in
wire attached to TB-7.
Diagnostics and troubleshooting
56
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...