3.
Is the sensor properly wired to the transmitter?
4.
Is a ground loop present? Refer to
Process pH grossly wrong or noisy
8.4.9
Simulating input currents - dissolved oxygen
To check the performance of the transmitter, use a decade box to simulate the current
from the oxygen sensor.
1.
Disconnect the anode and cathode leads from terminals 13 and 14 and connect a
decade box as shown in
It is not necessary to disconnect the RTD leads.
Simulate dissolved oxygen
Figure 8-11:
2.
Set the decade box to the resistance shown in the table.
Sensor
Polarizing voltage
Resistance
Expected current
499ADO
-675 mV
34 k
Ω
20 µA
499ATrDO
-800 mV
20 k
Ω
40 µA
Hx438 and Gx448
-675 mV
8.4 M
Ω
80 nA
3.
Note the sensor current. To view the sensor current:
a. Go to the main display and press
DIAG
.
b. Press
NEXT
.
SEnSor Cur appears in the display.
c. Press
ENTER
.
The display shows the sensor current.
d. Note the units.
µA is microamps; nA is nanoamps.
Diagnostics and troubleshooting
112
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...