Voltage (mV)
pH (at 25 °)
295.8
2.00
177.5
4.00
59.2
6.00
59.2
8.00
177.5
10.00
295.8
12.00
Preamplifier in a junction box
The procedure is the same as described in
. Keep the connection
between the analyzer and the junction box in place. Disconnect the sensor at the sensor
side of the junction box and connect the voltage source to the sensor side of the junction
box.
Preamplifier in sensor
The preamplifier in the sensor converts the high impedance signal into a low impedance
signal without amplifying it. To simulate pH values, follow the procedure in
8.4.12
Simulating temperature
General information about simulating temperature
The transmitter accepts either a Pt100 RTD (used in pH, 499ADO, 499ATrDO, 499ACL-01,
499ACL-02, and 499AOZ sensors) or a 22k NTC thermistor (used in Hx438 and Gx448 DO
sensors and most steam-sterilizable sensors from other manufacturers). The Pt100 RTD
has a three-wire configuration. See
Figure 8-14
. The thermistor has a two-wire
configuration.
Diagnostics and troubleshooting
116
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...