183
Reference Manual
00809-0100-4021, Rev GC
Section 6: Certified Safety Instrumented System (Safety-Certified)
April 2014
Certified safety instrumented system (safety-certified)
Abbreviated proof test
Conducting the abbreviated proof test detects approximately 63% of transmitter DU failures,
and approximately 90% of temperature sensor(s) DU failures, not detected by the 3144P
safety-certified automatic diagnostics, for a typical overall assembly coverage of 67%.
1.
Using Loop Test, enter the milliampere value representing a high alarm state.
2.
Check the reference meter to verify the mA output corresponds to the entered value.
3.
Using Loop Test, enter the milliampere value representing a low alarm state.
4.
Check the reference meter to verify the mA output corresponds to the entered value.
5.
Use a Field Communicator to view detailed device status to ensure no alarms or
warnings are present in the transmitter.
6.
Check that sensor value(s) are reasonable in comparison to a basic process control
system (BPCS) value.
7.
Document the test results per the plant’s requirements.
Extended proof test
Conducting the extended proof test, which includes the abbreviated proof test, detects
approximately 96% of transmitter DU failures and approximately 99% of temperature sensor(s)
DU failures, not detected by the 3144P safety-certified automatic diagnostics, for a typical
overall assembly coverage of 96%.
1.
Execute the abbreviated proof test.
2.
Perform a minimum two point sensor verification check. If two sensors are used, repeat
for each sensor. If calibration is required for the installation, it may be done in
conjunction with this verification.
3.
Verify that the housing temperature value is reasonable.
4.
Document the test results per the plant’s requirements.
Proof test intervals are based on typical sensor failure rates from the
Electrical and Mechanical
Component Reliability Handbook
Second Edition, exida.com, 2008. A low stress environment
without extension wire is assumed, with 30% of SIL 2 PFDavg limit budgeted for the transmitter
and sensor element. See the FMEDA report for additional details or references.
Table 6-1. Proof test interval guideline
Sensors
SFF
Abbreviated proof
test
Extended proof
test
Notes
4-wire RTD
90.8%
10 years
10 years
T/C
92.0%
10 years
10 years
Dual T/C
92.9%
10 years
10 years
Using U3 Drift Alert & HotBackup
Dual 3-wire RTD
92.5%
10 years
10 years
Using U3 Drift Alert & HotBackup
T/C & 3-wire RTD
91.2%
10 years
10 years
Using U3 Drift Alert & HotBackup
Содержание Rosemount
Страница 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Страница 2: ......
Страница 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Страница 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Страница 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Страница 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Страница 233: ...227 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 234: ...228 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 235: ...229 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 237: ...231 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 238: ...232 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 244: ...238 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Страница 251: ......