171
Reference Manual
00809-0100-4021, Rev GC
Section 4: F
OUNDATION
fieldbus Configuration
April 2014
Foundation™ fieldbus configuration
4.11.1
F
OUNDATION
fieldbus
If a malfunction is suspected despite the absence of a diagnostics message, follow the
procedures described in
to verify that transmitter hardware and process connections
are in good working order. Under each of the symptoms, specific suggestions for solving
problems are offered. Always deal with the most likely and easiest-to-check conditions first.
Table 4-12. F
OUNDATION
fieldbus Troubleshooting
Symptom
Potential source
Corrective action
Transmitter does not
Communicate with the
Configuration Interface
Loop Wiring
•Check for adequate voltage to the transmitter. The transmitter requires
between 9.0 and 32.0 V at the terminals to operate and provide
complete functionality
•Check for intermittent wire shorts, open circuits, and multiple grounds.
Network
parameters
•See
“Failure mode switch” on page 188
.
High Output
Sensor Input
Failure or
Connection
•Enter the transmitter test mode to isolate a sensor failure.
•Check for a sensor open circuit.
•Check the process variable to see if it is out of range.
Loop Wiring
•Check for dirty or defective terminals, interconnecting pins, or
receptacles.
Electronics Module
•Enter the transmitter test mode to isolate a module failure.
•Check the sensor limits to ensure calibration adjustments are within the
sensor range.
Erratic Output
Loop Wiring
•Check for adequate voltage to the transmitter. The transmitter requires
between 9.0 and 32.0 V at the terminals to operate and provide
complete functionality.
•Check for intermittent wire shorts, open circuits, and multiple grounds.
Electronics Module
•Enter the transmitter test mode to isolate module failure.
Low Output or No
Output
Sensor Element
•Enter the transmitter test mode to isolate a sensor failure.
•Check the process variable to see if it is out of range.
Loop Wiring
•Check for adequate voltage to the transmitter. The transmitter requires
between 9.0 and 32.0 V at the terminals to operate and provide
complete functionality.
•Check for wire shorts and multiple grounds.
•Check the loop impedance.
•Check wire insulation to detect possible shorts to ground.
Electronics Module
•Check the sensor limits to ensure calibration adjustments are within the
sensor range.
•Enter the transmitter test mode to isolate an electronics module failure.
Содержание Rosemount
Страница 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Страница 2: ......
Страница 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Страница 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Страница 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Страница 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Страница 233: ...227 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 234: ...228 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 235: ...229 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 237: ...231 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 238: ...232 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 244: ...238 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Страница 251: ......