Reference Manual
00809-0100-4021, Rev GC
Section 4: F
OUNDATION
fieldbus Configuration
April 2014
Foundation™ fieldbus configuration
Sensor calibration, lower and upper trim methods
In order to calibrate the transmitter, run the Lower and Upper Trim Methods. If your system does
not support methods, manually configure the Transducer Block parameters listed below.
1.
Set MODE_BLK.TARGET_X to OOS.
2.
Set SENSOR_CAL_METHOD_X to User Trim.
3.
Set CAL_UNIT_X to supported engineering units in the transducer block.
4.
Apply temperature that corresponds to the lower calibration point and allow the
temperature to stabilize. The temperature must be between the range limits defined
in PRIMRY_VALUE_RANGE_X.
5.
Set values of CAL_POINT_LO_X to correspond to the temperature applied by the
sensor.
6.
Apply temperature, temperature corresponding to the upper calibration.
7.
Allow temperature to stabilize.
8.
Set CAL_POINT_HI_X.
Note
CAL_POINT_HI_X must be within PRIMARY_VALUE_RANGE_X and greater than
CAL_POIN CAL_MIN_SPAN_X.
9.
Set SENSOR_CAL_DATE_X to the current date.
10.
Set SENSOR_CAL_WHO_X to the person responsible for the calibration.
11.
Set SENSOR _CAL_LOC_X to the calibration location.
12.
Set MODE_BLK.TARGET_X to AUTO.
Note
If trim fails the transmitter will automatically revert to factory trim.
Excessive correction or sensor failure could cause device status to read “calibration error.” To
clear this, trim the transmitter.
Recall factory trim
To recall a factory trim on the transmitter, run the Recall Factory Trim. If your system does not
support methods, manually configure the transducer block parameters listed below.
1.
Set MODE_BLK.TARGET_X to OOS.
2.
Set SENSOR_CAL_METHOD_X to Factory Trim.
3.
Set SET_FACTORY_TRIM_X to Recall.
4.
Set SENSOR_CAL_DATE_X to the current date.
5.
Set SENSOR_CAL_WHO_X to the person responsible for the calibration.
Содержание Rosemount
Страница 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Страница 2: ......
Страница 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Страница 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Страница 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Страница 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Страница 233: ...227 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 234: ...228 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 235: ...229 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 237: ...231 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 238: ...232 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 244: ...238 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Страница 251: ......