WARNING
ELECTRIC SHOCK
Fuses
MUST
only be replaced with the fuses of the same type and rating as those
being removed (6.3A 250 VAC ceramic cartridge fuse, 5x20 mm, Speed F. IEC
60127 compliant). Failure to do so may result in personnel injury and or equipment
damage.
4. Carefully lever open the fuse cover (B) to gain access to the fuses (C).
Figure 7-17: Power Socket Fuses
A. Power socket
B. Fuse cover/fuse holder
C. Fuse
5. Remove the fuse (C) from the holder (B).
6. Fit a new fuse (C), of the same type and rating, into the fuse holder (B) (6.3A 250
VAC ceramic cartridge fuse, 5x20 mm, Speed F. IEC 60127 compliant).
7. Repeat for the second fuse.
8. Fully push the fuse holder (B) back into position in the power socket (A).
9. Reconnect the external power supply to the power socket on the back panel.
10. Start up the analyzer by turning the
ON/OFF
switch on the back panel to the
ON
position.
11. The analyzer will automatically begin the startup process.
Troubleshooting and diagnostics
Reference Manual
January 2020
00809-0100-4244
84
Rosemount CT4400
Содержание Rosemount CT4400
Страница 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Страница 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Страница 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Страница 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Страница 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Страница 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Страница 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Страница 105: ......