Figure 2-4: Gas Inlet and Outlet Connections
A. Rear panel
B. Sample gas outlet
C. Sample gas inlet
2.3
Connecting the electrical/electronic inputs and
outputs
This section discusses making the electrical connections to the analyzer.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (
Death or personal injury may result if this is not observed.
Install
Reference Manual
January 2020
00809-0100-4244
26