3. To abort the calibration process, press
or
, otherwise the analyzer will
perform the
SPAN CALIBRATION
, either after waiting for the purge time, or when
or
is pressed, whichever is sooner. The result of the calibration will then be
displayed, as below.
Figure 6-18: Span Calibration Result
4. Press the appropriate button.
Option
Description
To return to the
Gas Service
Screen.
To select the
Next Gas
from the
Select
Gas
screen.
To return to the
Main Menu
.
To return to the
Gas Sensor Main
screen
(
Gas calibration
Reference Manual
January 2020
00809-0100-4244
60
Rosemount CT4400