2. Allow the concentration to stabilize.
Figure 6-7: Calibrate Zero for NH3
This screen gives a reading of the concentration of the selected gas that is present
as an impurity in the nitrogen calibration gas, as measured by the analyzer.
In the example shown the gas being measured is NH
3
(ammonia), and the analyzer
has detected a concentration of 0.510 ppm (vol).
3. To abort the calibration process, press
▲
or
▼
, otherwise the analyzer will perform
the
ZERO CALIBRATION
, either after waiting for the purge time, or when
►
or
is
pressed, whichever is sooner. The result of the calibration will then be displayed, as
below.
Figure 6-8: Result
4. Press the appropriate button
Option
Description
To return to the
Gas Service Screen
.
Gas calibration
Reference Manual
January 2020
00809-0100-4244
54
Rosemount CT4400